Model Number BEA28-80/I16-40
Medical Device Problem Codes Use of Device Problem (1670); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device remains implanted in the patient; therefore, it will not be returned for evaluation.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Event or Problem Description
The patient was initially implanted with an afx2 bifurcated stent graft and a suprarenal extension to treat an abdominal aortic aneurysm.Approximately 4 years post initial procedure the patient was identified with a type 1b endoleak during a routine follow up.The patient was reported as stable and the physician is planning a re-intervention.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured, and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because the device remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type ib endoleak is unconfirmed and the additional endovascular procedure is confirmed.This is moderately consistent with the reported adverse event/incident.Though the type ib endoleak could not be confirmed, the right common iliac was off label measuring at 2.8cm and should be 10-23mm.Additionally, there was the absence of an abdominal aortic aneurysm which is off label.A definitive root cause for the reported adverse event/incident could not be determined; however, the off-label use could have been a contributing factor.Device and procedure related harms for this event could not be determined.The final patient status was reported as stable.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.H6: health effect - impact code ¿ remove 4614.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
 
Event or Problem Description
The patient was initially implanted with an afx2 bifurcated stent graft and a suprarenal extension to treat an abdominal aortic aneurysm.Approximately 4 years post initial procedure the patient was identified with a type 1b endoleak during a routine follow up.The patient was reported as stable and the physician is planning a re-intervention.After the initial report, additional information was received, reporting that reintervention was completed with the implant of an ovation ix extender.The type 1b endoleak was successfully sealed, and there were no adverse complications.The final patient status was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key15297899
Report Number2031527-2022-00212
Device Sequence Number5839048
Product Code MIH
UDI-Device Identifier00818009014467
UDI-Public(01)00818009014467(17)181013
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/13/2018
Device Model NumberBEA28-80/I16-40
Device Lot Number1749754-010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer09/30/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/01/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL SERIAL# (B)(6)
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age75 YR
Patient SexMale
-
-