| Model Number |
MON-000005 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Chest Pain (1776); Pneumothorax (2012)
|
| Date of Event |
06/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This event was reported from the transbronchial biopsy assisted by robot guidance in the evaluation of tumors of the lung (target) study 19-br-0001 (nct04182815).The device was not returned for evaluation.There was no allegation of device failure.Risk of pneumothorax is documented in 105-000299-01, rev f, monarch bronch risk management report and is deemed to be acceptable.Based on the information available for this complaint, the root cause of the reported event is related to a known inherent risk of the device and procedure as documented in the labeling.Serial number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a history of lupus and a recent bronchoscopy for a biopsy of the left lung nodule returned to the hospital the following morning for pain in the left chest and requested a chest x-ray for further evaluation.The patient denies shortness of breath but reports increased pain with deep breathing.The patient did have a post procedure x-ray done following the procedure which showed no pneumothorax.The patients pain started on the way home from the procedure the previous day.A thora vent was placed into the left chest.Suctioning was used for about an hour.A repeat x-ray at that time showed near complete resolution of the pneumothorax on the left.The patient continued to feel well and was not hospitalized.The patient returned the following day to have the thora vent removed and reported tolerated the device well, no shortness of breath as well as no significant left chest pain.
|
| |
|
Additional Manufacturer Narrative
|
|
The follow-up report is be submitted to relay additional information.The following sections were updated: there were no device malfunctions reported from the procedure.A device history record (dhr) review was performed for system 110024 and there were no non-conformances related to the pneumothorax reported in this case.Based on the information available for this complaint, the root cause of the reported event is related to a known inherent risk of the device and procedure as documented in the labeling.No corrective action is required at this time.The reported event will continue to be monitored through regular complaint trend review.The follow-up report is be submitted to relay corrected information.The following sections were updated: g3, g4 g3: 09/12/2022.G4: k173760.
|
| |
|
Search Alerts/Recalls
|