| Model Number |
23AGFN-756 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It as reported that on (b)(6) 2022, a 23mm sjm regent heart valve with flex cuff was selected for procedure.It was noted that during procedure a leaflet of the regent valve become dislodged.The decision was made to replace the 23mm sjm regent heart valve with flex cuff with another one of the same size.However, it was noted that there was a 60 minute delay in procedure because the replacement valve had to be delivered from a different hospital.The patient was discharged at the time of report.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of a dislodged leaflet was confirmed.One leaflet was unfractured and dislodged from the orifice.No other damages were noted on the valve.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The root cause of the leaflet dislodgement could not be conclusively determined.
|
| |
|