MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO NYLON; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
NV-7-30-HELIX |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received a report that two concerto coils experienced resistance in the proximal part of the microcatheter and one prematurely detached during removal.The patient was undergoing treatment for gi bleed embolization.The patient's blood flow was normal, and their vessel tortuosity was moderate.It was reported that first 5x15 coil was deployed as an anchor.Another 5x20 was used, but it jammed immediately in the microcatheter.They upsized and used 2 7x30 coils successfully.A third 7x30 coil jammed in the microcatheter and detached when the doctor pulled the introducer sheath back to see if the coil was in the microcatheter or jamming in the hub. the coil was pulled out of the catheter hub with no further medical/surgical intervention required.All coils were dipped in saline and then sheathed before implanting.When the coils jammed they were resheathed and re-dipped in the saline.Then they were tried again but jammed.The pushwire was not bend or broken, the physician did not reposition the coil, no detachment attempts had been made, the physician did not rotate the delivery pusher, and a continuous flush was not administered. the coils had pushed out of the introducer sheath smoothly, and there was no damage to the coils or catheter. the coil was replaced, and the patient did not experience any injury or complications.The devices were prepared according to the instructions for use (ifu).Ancillary devices include a boston scientific renegade high flow microcatheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the concerto coil was returned for analysis.The concerto pusher actuator interface (ai) and hypotube break indicator (hbi) were found intact.The concerto coil was returned within its introducer sheath.The implant coil was found detached from the pusher.The detached implant coil was not returned.The pusher was removed from within the introducer sheath.The pusher was found bent from ~20.0cm to ~8.0cm from the pusher distal end.The pusher release wire coin was found against the lumen stop and the shield coil intact was intact.Based on the device analysis and reported information, the customer¿s ¿coil resistance/stuck in catheter¿ and ¿premature detachment¿ reports were confirmed.Damage to the concerto pusher can occur if the device is advanced against resistance.The damage to the pusher can cause the release wire coin to temporarily pull back from the lumen stop detaching the implant coil.Possible contributing factors to ¿coil resistance¿ include using an incompatible device for delivery, tortuous anatomy, failure to maintain a continuous flush or failure to hydrate the concerto coil before use.The (boston scientific) renegade hi-flo micro catheter was not returned.The renegade hi-flo micro catheter has a labeled id (inner diameter) of 0.027¿, as per an online source.Per the concerto ifu (instructions for use): concerto nylon detachable coils with diameters of 5mm to 10mm should be delivered through a microcatheter with a minimum inside diameter of 0.021¿.Therefore, the renegade micro catheter was found to be compatible for use with the concerto coil.The patient¿s vessel tortuosity was ¿moderate¿ and the concerto coil was prepared before use.In this event it is likely the failure to maintain a continuous flush contributed to the event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|