| Model Number |
VTICM5_13.2 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm, vticm5_13.2, -10.0/4.0/095 (sphere/cylinder/axis) implantable collamer lens was torn.The reporter stated " either defect or during load ".No patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
Expiration date: 12/31/2023.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5- per updated information the lens was noted to be damaged after loading and viewing in cartridge under microscope.A backup lens was successfully implanted.Claim # (b)(4).
|
| |
|
Search Alerts/Recalls
|