• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AROA BIOSURGERY LTD. OVITEX REINFORCED TISSUE MATRIX; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AROA BIOSURGERY LTD. OVITEX REINFORCED TISSUE MATRIX; SURGICAL MESH Back to Search Results
Model Number F10248-1020P
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 07/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Intraperitoneal placement of surgical mesh carries a known risk of adhesions.Such adhesions, whether device or surgery related, may lead to small bowel obstruction.It cannot be definitively determined whether the device contributed to this event.No device problem was found from review of the dhr.The devices in ert-21f12 were produced as per established manufacturing procedures.All process specifications including package labelling requirements were met.Ckage labelling requirements were met.
 
Event or Problem Description
A small bowel obstruction was reported after the use of ovitex 2s permanent in open ipom sugarbaker parastomal hernia repair.This required reoperation at which time there were noted loops of small bowel attached to the device and to the colon on the opposite side.The surgeon took down adhesions and corrected tears as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVITEX REINFORCED TISSUE MATRIX
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
AROA BIOSURGERY LTD.
2 kingsford smith place
airport oaks, auckland 2022
NZ  2022
Manufacturer (Section G)
AROA BIOSURGERY LTD.
2 kingsford smith place
airport oaks, auckland 2022
NZ   2022
Manufacturer Contact
tina o'brien
2 kingsford smith place
airport oaks, auckland 2022
NZ   2022
MDR Report Key15299217
Report Number3007321028-2022-00030
Device Sequence Number7092044
Product Code FTM
UDI-Device Identifier09421904065178
UDI-Public09421904065178
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K153632
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberF10248-1020P
Device Catalogue NumberF10248-1020P
Device Lot NumberERT-21F12
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2022
Initial Report FDA Received Date08/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-