| Model Number |
2D72PL70X |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Anaphylactic Shock (1703)
|
| Date of Event |
07/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During placement of an implantable catheter for chemotherapy, it was reported that from the clinician¿s sterile nitrile gloves, the patient developed a sensation of generalized pruritus, generalized rash with appearance of hives, feeling of dyspnea, dizziness with tendency to worsen.Patient was sent to the emergency room for monitoring of condition and anaphylaxis.Patient required 02 therapy to raise saturation from 85-87% to 94%.40mg of defidramine iv and 100mg of hydrocortisone were administered to the patient.The patient was reported as stable after treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint was forwarded to the manufacturing facility where it is currently still under investigation.A follow-up report will be filed once the results have been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Supplemental report is being filed since the results of investigation are available.The device history record of the complaint lot was reviewed.The lot was inspected and released in compliance with all requirements.The sample was not returned for evaluation but only wallet paper photo provided.Investigation found no change in material and process that may cause the reported incident in complaint lot.Reference to tsr#90077 test report that passed primary skin irritation (psi) testing per iso10993-10:2010 and the product also passed protein test both of astm d5712-15 and en455-3:2015.Pi classic gloves have passed a series of tests prescribed by regulatory agencies for the intended use.Therefore, the root cause could not determined for this case.We will continue to monitor trends for this product.
|
| |
|
Search Alerts/Recalls
|