• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 07/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not returned; therefore, a failure analysis of the complaint device cannot be completed.A review of the device record and complaint trending for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from a patient's left eye because patient is bothered by loss of contrast and poor visual acuity quality.Pl-0.50x30 20/20.Pre-op visual acuity: 20-30 -1.The patient was adament about removing the lens, very poor vision for distance.The incision was enlarged; however, no other surgical interventions were required.The explanted iol was replaced with a non-johnson and johnson iol.Post-explant post-op visual acuity: 20/20.No other information was provided.
 
Additional Manufacturer Narrative
D9.Device available for evaluation? yes.Returned to manufacturer on: sep.19, 2022.Section h3.Device evaluated manufacturer? yes.Device evaluation: the complaint lens was received with a cut on the optic body, which is consistent with a lens that was handled during explant.No further defects that could contribute to visual issues were observed.Based on the return condition of the lens no further product evaluation could be performed.The complaint issues ¿hm-visual disturbance¿ and ¿hm-sub-optimal results¿ were not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: as a result of the investigation there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15299348
Report Number3012236936-2022-02092
Device Sequence Number15238590
Product Code HQL
UDI-Device Identifier05050474750654
UDI-Public(01)05050474750654(17)241118
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/18/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer10/10/2022
Initial Report FDA Received Date08/25/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight75 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-