• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC LITE METER FREESTYLE; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC LITE METER FREESTYLE; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 70805-70
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An extended investigation has been conducted, which involved a manufacturing review (including 5 years of device history records (dhrs)), previous complaint and corrective and preventive action (capa) investigations conducted for blank screen issues, process failure mode effects analyses (pfmeas), design controls, and design failure mode effects analyses (dfmeas), risk management reports, risk evaluations, and label copy.The investigation did not identify any indication that the product did not meet specification.If the product is returned, an investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
The returned meter was investigated with retain test strips.Performed visual inspection on returned meter and no issues were observed.Meter powered on with button depression and test strip insertion.No new issues were observed.Blank screen was not observed.The complaint was not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LITE METER FREESTYLE
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key15299900
Report Number2954323-2022-31167
Device Sequence Number7307258
Product Code NBW
UDI-Device Identifier00699073000019
UDI-Public00699073000019
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K092602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number70805-70
Device Catalogue Number70804
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer09/27/2022
Initial Report FDA Received Date08/26/2022
Supplement Report FDA Received Date09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age32 YR
Patient SexUnknown
Patient Weight75 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-