STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6020-0435
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body Reaction (1868)
Date of Event 08/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient is a 68 year old male who underwent surgery on (b)(6) 2010 for a left total hip arthroplasty, presented with pain and x-rays showed a femoral head dissociation from the trunnion of the stem as well as the trunnion fracturing off the neck of the body of the stem and floating free in the joint space.The original stryker implants implanted were: accolade tmzf size 4x132 stem, a 36-5 v40 cocr femoral head, a size 54 trident ha hemispherical acetabular shell, and two screws, and size 36 e neutral x3 polyethylene liner.No implant record available for the first surgery.The patient was brought into the operating room, a revision hip was performed on the left side, the trunnion was fractured from the stem, the head was free in the acetabulum causing impingement and metallosis.There was adverse local tissue reactions from the trunnionosis however the adductors were in tact.Another company's revision stem was used, the trident cup was retained and a new 36mm size e 10 degree liner was put into the cup.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE TMZF HIP STEM #4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15300188
Report Number0002249697-2022-01244
Device Sequence Number14070369
Product Code LPH
UDI-Device Identifier04546540510563
UDI-Public04546540510563
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2011
Device Model Number6020-0435
Device Catalogue Number6020-0435
Device Lot Number19356801
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/02/2022
Initial Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/10/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexMale
-
-