A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: analysis of the returned device identified: underweight, opening on posterior and red particles in the shell.A visual and microscopic analysis was performed which identified: sharp opening on posterior, stress marks and crease fold.A dimension measurement in the shell was performed which identify the thickness within specification.Base on the device analysis the final assessment is: a sharp pattern on posterior of opening device assessed as a surgical impact opening, for the stress mark in the shell.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: patient's concern with the product.
|