| Model Number |
10-85-00 |
| Medical Device Problem Code |
Pumping Stopped (1503)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no patient involvement.Livanova deutschland manufactures the s5 double head pump.The incident occurred in milano, italy.A livanova field service representative was dispatched to the facility and replaced the motor control board as precautionary measure.Subsequent functional verification testing was completed without further issues.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova received report that a s5 double roller pump gave a motor control failure error message during priming.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H.10: the serial read-out of the pump (real time device parameters and setting recording file) was analyzed and confirmed that on the date of the event the reported error code was stored in the pump microcontroller.The reported error code can be solely related to a fault of the motor control board of the pump which was replaced as per initial report.Based on collected information and service findings, the root cause of the reported event is traced back to an intermittent malfunction of the motor control board.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|