• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM LIGHTNING 7; QEW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM LIGHTNING 7; QEW Back to Search Results
Catalog Number LITNG7XTORQ130
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a thrombectomy procedure for a femoropopliteal bypass using an indigo system aspiration catheter 7 (cat7) and a non-penumbra sheath.It was noted that a stent was placed inside the patient prior to the procedure.During the procedure, the physician advanced the cat7 through the stent, removed the guidewire and completed couple of passes using the cat7.Subsequently, while attempting to remove the cat7 out of the patient, the physician experienced resistance and the cat7 became stuck.The physician attempted to pull the cat7; however, the cat7 broke in half.The physician then used a pair of hemostats and removed the broken end of the cat7.The procedure was completed by placing a stent.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned cat7 confirmed a fracture on the mid-shaft.If the cat7 is retracted against resistance, damage such as this may occur.The root cause of the resistance could not be determined.Further evaluation of the device revealed ovalizations and kinks throughout the length of the catheter.This damage was incidental to the complaint and likely occurred during removal from the patient with hemostats and during packaging for the device return.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM LIGHTNING 7
Common Device Name
QEW
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15302697
Report Number3005168196-2022-00393
Device Sequence Number5339557
Product Code QEW
UDI-Device Identifier00815948022584
UDI-Public815948022584
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K210323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/26/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/07/2025
Device Catalogue NumberLITNG7XTORQ130
Device Lot NumberH12519
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2022
Initial Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-