| Model Number |
VTICM5_13.2 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The reporter indicated that on (b)(6) 2022 a 13.2mm, vticm5_13.2, -8.00/2.0/090 (sphere/cylinder/axis) implantable collamer lens was implanted, removed and replaced intraoperatively due to the doctor noticing a missing haptic after implanting.It was reported that the problem resolved.Cause of event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Expiration date: 12/31/2023.Lens work order search: no similar complaint type events reported for units within the same lot.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: the lens was returned a vial with liquid.Visual inspection found the haptic torn.Claim# (b)(4).
|
| |
|