Model Number 500-301
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Unintended Radiation Exposure (4565)
Date of Event 07/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient''s date of birth unk.Patient''s weight unk.Patient''s ethnicity/race unk.Relevant tests/laboratory data unk.Other relevant history unk.The device was discarded, thus no investigation could be completed.The device was discarded, thus the reported failure could not be established.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event or Problem Description
A lead extraction procedure commenced to remove a right ventricular (rv) lead due to non-function.A spectranetics lead locking device (lld) was inserted into the rv lead to provide traction.The physician chose a spectranetics 12f glidelight laser sheath to assist with extraction attempt.When the glidelight was removed from the patient, a split in the catheter was noted.A spectranetics 11f tightrail rotating dilator sheath was used to successfully remove the rv lead with no reported patient harm.This event is being reported for unintended radiation exposure, potential for harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS 12F GLIDELIGHT LASER SHEATH
Common Device Name
DEVICE, REMOVAL, PACEMAKER ELECTRODE, PERCUTANEOUS
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key15303063
Report Number1721279-2022-00155
Device Sequence Number6484082
Product Code MFA
UDI-Device Identifier00813132020323
UDI-Public(01)00813132020323(17)240412(10)FGA22D07A
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P960042 S070
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/12/2024
Device Model Number500-301
Device Catalogue Number500-301
Device Lot NumberFGA22D07A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2022
Initial Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MEDTRONIC 6935 RV ICD LEAD; SPECTRANETICS 11F TIGHTRAIL DILATOR SHEATH; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; SPECTRANETICS LEAD LOCKING DEVICE
Patient Age79 YR
Patient SexMale
-
-