| Catalog Number |
07154003001 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they have an issue with the cobas infinity software.For one patient, a test result was being saved erroneously in the cobas infinity software.A urine glucose value was saved as a 24 hour urine glucose value.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined the issue was due to a misconfiguration in the test mapping along with a software error in cobas infinity.This allowed for tests with a non-corresponding tube type and non-corresponding result to be added to an order.
|
| |
|