C.R. BARD, INC. (BASD) -3006260740 HUBER PLUS 20G X 1" NEEDLELESS Y SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refq1860 showed twelve other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that nurse opened a huber plus and checked patency.When flushed with heparinized saline from the main hub, there was leaking from y-connector hub.Nurse kept the defective needle and opened a new box of huber plus to continue treatment.Patient status/ intervention: cancer patient is fine.
 
Event or Problem Description
It was reported that nurse opened a huber plus and checked patency.When flushed with heparinized saline from the main hub, there was leaking from y-connector hub.Nurse kept the defective needle and opened a new box of huber plus to continue treatment.Patient status/ intervention: cancer patient is fine.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of a y-site leak is confirmed but the exact cause remains unknown.One video sample of a huber plus infusion set was provided for evaluation.The valved y-site is shown in the video with a clear fluid being infused from the proximal luer connector.A jet of fluid is observed to leak from the y-site valve region.The characteristics of the leak location and condition of the device could not be inspected from the video provided.Based on the information provided, possible contributing factors include hub and y-site valve damage.Since a leak was observed to be present, the complaint is confirmed but the exact cause cannot be determined.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUBER PLUS 20G X 1" NEEDLELESS Y SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15303464
Report Number3006260740-2022-03353
Device Sequence Number5287789
Product Code FPA
UDI-Device Identifier00801741065569
UDI-Public(01)00801741065569
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K993848
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number012001NY
Device Lot NumberREFQ1860
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer09/20/2022
Initial Report FDA Received Date08/26/2022
Supplement Report FDA Received Date09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-