• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED STERILIZATION PRODUCTS STERRAD® NX STERILIZER; STERRAD® EQUIPMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED STERILIZATION PRODUCTS STERRAD® NX STERILIZER; STERRAD® EQUIPMENT Back to Search Results
Model Number 10033
Medical Device Problem Code Environmental Particulates (2930)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported an event of a ¿smoke¿ or haze emitting from the sterrad® nx sterilizer.There was no report of any injuries or human reactions.A field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
 
Additional Manufacturer Narrative
Initial reporter phone number: (b)(6).A field service engineer was dispatched to the customer site.The oil mist filter was replaced to resolve the smoke/haze issue.Unit meets specifications and was returned to service.The device history record (dhr) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by advanced sterilization products, or its employees that the report constitutes an admission that the product, advanced sterilization products, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Asp complaint ref #: (b)(4).
 
Additional Manufacturer Narrative
Correction: block b5 in the initial report contained incorrect information - the customer reported a separate issue for the sterrad® nx, and while on site to assess the unit, the field service engineer observed the smoke/haze issue.H3: asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the smoke/haze issue, and system risk analysis (sra).Trending analysis of the smoke/haze issue for the sterrad® nx unit was reviewed for the prior six months from the event date and no significant trend was observed.Review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." no parts were available for further analysis.The assignable cause of the smoke/haze is likely due to the oil mist filter.The field service engineer replaced the part and confirmed the sterrad® nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD® NX STERILIZER
Common Device Name
STERRAD® EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer (Section G)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer Contact
gabriela mclellan
33 technology drive
irvine, CA 92618
9495030264
MDR Report Key15303559
Report Number2084725-2022-00256
Device Sequence Number17732594
Product Code MLR
UDI-Device Identifier10705037014132
UDI-Public10705037014132
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K042116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10033
Device Catalogue Number10033
Device Lot NumberN/A
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2022
Supplement Date Received by Manufacturer08/28/2022
Initial Report FDA Received Date08/26/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-