| Model Number |
MT22949 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the receiver display was frozen.No product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).B5: describe event or problem - additional.D9: device availability - additional.H2: additional information/device evaluation.H3: device evaluated by manufacturer - additional.H6: adverse event problem - additional.
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An external visual inspection was performed and passed.A receiver charge test was not performed due to the receiver perpetually rebooting.A receiver boot test was performed and failed due to the receiver perpetually rebooting.Functional tests were not performed due to the receiver perpetually rebooting.An internal visual inspection was performed and failed due to moisture damage.Pictures were taken.The receiver log was not downloaded for review due to the receiver perpetually rebooting.The allegation was undetermined.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|