• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JJGC S.A. CM DRIVE ACQUA IMPL TI 5.0X10; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JJGC S.A. CM DRIVE ACQUA IMPL TI 5.0X10; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Model Number 140.685
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.  the loss of an endosseous dental implant after successful osseointegration and restoration is a known inherent risk of the procedure.Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia 2.Absence of a recurrent peri-implant infection with suppuration 3.Absence of implant mobility 4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2019 in ada 19.On (b)(6) 2021, loss of osseointegration was verified.The device was forwarded to the manufacturer.There were no patient operative or post-operative complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM DRIVE ACQUA IMPL TI 5.0X10
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
Manufacturer (Section G)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR   81270-200
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key15304824
Report Number3008261720-2022-11890
Device Sequence Number6485052
Product Code DZE
UDI-Device Identifier07899878016861
UDI-Public07899878016861
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K133592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/11/2020
Device Model Number140.685
Device Catalogue Number140.685
Device Lot Number800341944
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/26/2022
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexMale
-
-