No device was returned for evaluation; however, two (2) anteroposterior radiographs provided, three (3) photographs of the screwdriver, and an onsite representative were able to confirm the reported tip fracture.Examination of the photographs confirmed approximately 5 mm of the tip had fractured and the residual fingers are slightly deformed, suggestive of excessive force.Based on the information obtained, the root cause of the reported tip fracture is unknown; however, excessive or off-angled force may be a likely cause or potential contributing factor.No additional investigation can be completed at this time.Labeling review: "pre-operative warnings.¿implants and instruments should be protected during storage and from corrosive environments.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." "pre-operative warnings.The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury." "intra-operative warnings.The physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted." device not returned to manufacturer.
|