| Model Number |
133S-MV-14-T |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation" against right device.The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation" against right device.The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation" against right device.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed a puncture opening assessed as to surgical damage like.Observed an opening sharp assessed as unidentified (tear) opening (shell thickness was within specification).In the sature tab observed a striated assessed as to surgical damage and missing piece of the shell assessed as to inconclusive.Additional observations: crease flat observed in the device.Wear abrasion observed in the surface of the shell.No further actions are required as the device was implanted.Additional, changed, and/or corrected data: d.9., h.2., h.3., h.6.
|
| |
|