Model Number 48630W-04
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 07/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, the cause of the customer reported failure mode and intra-operative complication cannot be determined.Isi has requested the stapler 30 white reload to be returned for failure analysis evaluation, but the product has not been received as of the date of this report.A follow-up mdr will be submitted if additional information is obtained.A system log review was performed for this procedure: no relevant errors were observed during this procedure.The stapler logs for this event were reviewed by an isi failure analysis engineer (fae): per the fae, the logs show curved-tip stapler 30 instrument (part #470530-08, lot #t10210104-0042) was installed 6 times and fired 4 reloads (all white).All firings were completed per the logs.There were no incomplete clamps by this instrument.The logs show an e-stop press occurred roughly 1 hour after the 4th firing, and the system detected use of the stapler release kit (srk) on this instrument.It is not clear what prompted the user to use the srk.This complaint is being reported due to the following conclusion: during a da vinci-assisted pulmonary lobectomy procedure, it was alleged that staples were not deployed when a white stapler 30 reload was fired with a curved-tip stapler 30 instrument.As a result, the patient lost 4,200 ml of blood and the procedure was converted to open surgery to control the bleeding.The cause of the customer reported failure mode and intra-operative complication is unknown.
 
Event or Problem Description
Intuitive surgical inc.(isi) received fda voluntary report with mdr report #mw5110971 stating: ¿pt scheduled for robotic-assisted left upper lobectomy.Emergently converted to open procedure due to bleeding.It appeared that all of the staples didn¿t deploy and the blade was still exposed, causing vascular injury and exsanguination of 4200ml of blood 30mm stapler used ref#(b)(4), lot # t102101040042 with load model #48630w, lot m90210503.Fda safety report id# (b)(4).¿ intuitive surgical, inc.(isi) has attempted to obtain additional information.However, as of the date of this report, no further details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Common Device Name
STAPLER 30 RELOAD WHITE
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key15307278
Report Number2955842-2022-13735
Device Sequence Number6484621
Product Code GDW
UDI-Device Identifier10886874112868
UDI-Public(01)10886874112868(10)M90210503
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K152421
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Risk Manager
Remedial Action Other
Type of Report Initial
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48630W-04
Device Catalogue Number48630W
Device Lot NumberM90210503 641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2022
Initial Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-