| Model Number |
3100 B |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to vyaire medical device overheated and stopped oscillating during use on a patient on the 3100 b ventilator.At this time, there is no information regarding patient harm associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
"results of investigation: vyaire field service representative went onsite to evaluate the device.Technician troubleshot and tested unit.Left oscillator running for one hour with settings that were on the unit at arrival with no problem.Tested oscillator for another hour with the performance check settings and unit worked within manufacturer specifications.Unit is ready for use.".Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|