Model Number 83786
Medical Device Problem Codes Difficult to Remove (1528); Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.The main coil, the introducer sheath, the delivery wire, and a valve were returned.The main coil and the delivery wire were not interlocked.The introducer sheath was inspected and no damages were found.The twist lock was opened.The delivery wire was inspected, and no damages were found.The main coil was kinked, stretched, and the interlocking arm of the main coil was detached.
 
Event or Problem Description
Reportable based on device analysis received on (b)(6) 2022.It was reported that high resistance was noted while attempting to deploy the coil.The target lesion was located in the splenic artery.An 18mmx40cm and 20mmx40cm interlock-35 were selected for use.During the procedure, high resistance was encountered when deploying the 18mmx40cm coil with the help of a 5f catheter.It was then withdrawn and pulled back to the sheath and found that it could not be deployed.The entire catheter was removed.The coil was replaced, but the same issue occurred with the 20mmx40cm coil.The procedure was completed with another of same device.No complications were reported, and the patient was stable post procedure.However, device analysis revealed that the interlocking arm was detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK-35
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15307767
Report Number2124215-2022-31820
Device Sequence Number20423362
Product Code KRD
UDI-Device Identifier08714729795414
UDI-Public08714729795414
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133208
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/23/2024
Device Model Number83786
Device Catalogue Number83786
Device Lot Number0028443793
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/03/2022
Initial Report FDA Received Date08/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
Patient Weight67 KG
-
-