• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 7-8 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 7-8 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71453232
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, after a tka performed on (b)(6) 2013, on (b)(6) 2022 the post broke off the legion posterior stabilized high flex highly cross linked polyethylene size 7-8 11mm.This was addressed via revision surgery on (b)(6) 2022 to replace the insert.Current health status of the patient is unknown.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on a review of the information provided, the root cause of the breakage of the post on the insert could not be determined.However, the patient¿s body habitus of 305 pounds, age, activity level, and bone quality could not be ruled out as a possible contributory factor.According to the report, the legion posterior stabilized high flex highly cross-linked polyethylene size 7-8 11mm was replaced and no other parts were affected.It should also be noted the implant was implanted 9 years prior to the revision.The patient impact was instability, locking, pain, fractured post, the subsequent revision, and expected transient post-op convalescence period.According to the report, the patient¿s current health status is healthy.Therefore, no further clinical/medical assessment is warranted at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that the implant can break or become damaged as a result of strenuous activity or trauma, and has a finite expected service life and may need to be replaced in the future.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, size selected, surgical technique or postoperative care.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 7-8 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15307960
Report Number1020279-2022-03861
Device Sequence Number2797916
Product Code JWH
UDI-Device Identifier00885556035153
UDI-Public00885556035153
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/13/2020
Device Model Number71453232
Device Catalogue Number71453232
Device Lot Number10JM09692
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date08/28/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/16/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight138 KG
-
-