SMITH & NEPHEW ORTHOPAEDICS AG CS/CSL/GERADSCHAFT-PLUS EXTRACT.SCREW M6; OSTEOTOME Back to Search Results
Model Number 75002165
Medical Device Problem Codes Peeled/Delaminated (1454); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, during the set-up for a tha surgery, a cs/csl/geradschaft-plus extract screw m6 was cross threaded into the implant and the metal is beginning to peel off.As this was noticed during the set-up, the patient was not yet involved.
 
Additional Manufacturer Narrative
It was reported that, during the set-up for a tha surgery, a cs/csl/geradschaft-plus extract.Screw m6 was cross threaded into the implant and the metal is beginning to peel off.As this was noticed during sterilization activities, the patient was not involved.The device intended for use in treatment was returned for investigation.A product evaluation was performed, and it was found that the thread is damaged significantly and is not functional.So, the mentioned failure mode of material twisted can be confirmed.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.The performed complaint history review for the complaint device revealed 2 additional complaints reported for the same batch in scope with related failure mode, where the root cause is related to problems attributed to design constrains.A new design of the device has been released in order to reduce the reoccurrence of this issue.This version of the device will be monitored for similar issues.The performance of the device is nonetheless within the risks, which are anticipated in the risk management documentation of the product.No further escalation is required.A review of the risk management documentation verifies the failure mode and severity of the reported issue.A relationship between the reported event and the device can be confirmed, with the root cause related to design constrains.There is no indication that the reported devices failed to meet manufacturing specifications upon release for distribution.Considering the performed visual inspection, this device failure may arise from a ¿wear and tear issue¿.Regular wear and tear are known to contribute to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.The need for corrective action is not indicated.Nevertheless, smith+nephew will continue to monitor this device for similar issues.The returned device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS/CSL/GERADSCHAFT-PLUS EXTRACT.SCREW M6
Common Device Name
OSTEOTOME
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15307975
Report Number9613369-2022-00483
Device Sequence Number6181443
Product Code HWM
UDI-Device Identifier07611996056563
UDI-Public07611996056563
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75002165
Device Catalogue Number75002165
Device Lot NumberB71729
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date08/28/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/17/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-