• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL SDN. BHD. RUSCH CURVED REINFORCED CUFFED ET TUBE; PRESSURE CUFF TRACHEAL TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL SDN. BHD. RUSCH CURVED REINFORCED CUFFED ET TUBE; PRESSURE CUFF TRACHEAL TUBE Back to Search Results
Model Number IPN040963
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).
 
Event or Problem Description
It was reported that: "the doctor found cuff leakage during clinical setting before using on patient.No impact on patient.".
 
Event or Problem Description
It was reported that: "the doctor found cuff leakage during clinical setting before using on patient.No impact on patient.".
 
Additional Manufacturer Narrative
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.Functional testing was then performed and the cuff and pilot balloon were inflated to 25mbar using a calibrated endotest meter.No issues were encountered.The sample was then immersed in water for leak testing.No air bubbles were observed coming from the pilot balloon or cuff indicating no leakage.A device history record review was performed and no relevant findings were identified.The complaint of leakage could not be confirmed.No issues were found with the returned sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUSCH CURVED REINFORCED CUFFED ET TUBE
Common Device Name
PRESSURE CUFF TRACHEAL TUBE
Manufacturer (Section D)
TELEFLEX MEDICAL SDN. BHD.
perak, west malaysia
Manufacturer (Section G)
TELEFLEX MEDICAL SDN. BHD.
lot no : pt2577 jalen perusahaan
4 kamunting industrial estate
perak, west malaysia 34600
MY   34600
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15309724
Report Number8040412-2022-00257
Device Sequence Number2786488
Product Code BTR
UDI-Device Identifier14026704340900
UDI-Public14026704340900
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN040963
Device Catalogue Number103902-000040
Device Lot Number17LT23C
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer09/30/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-