On 6/16/2022, it was reported by a sales representative via email that on (b)(6) 2022, a patient underwent surgery for a trochanteric fracture and an arthrex short trochanteric nail was implanted.Twelve days post operative, while trying to stand from the wheelchair, patient felt like the nail had exploded out of the leg.X-rays were taken of the implants and while it is still unknown what happened, it seems like the locking ring was not able to hold the lag screw in place.As of now, the surgeon is not planning on a revision surgery and will follow up again with the patient in two weeks.Additional information provided on 08/08/2022, it was reported that another 4 week post op occurred on (b)(6) 2022 and x-rays showed that the nail had further backed out.A revision surgery took place on (b)(6) 2022, at this time the 1031-200 galileo® troch nail,1099-100 galileo lag screw, and 8001-036 captured cortical screw were explanted.The surgeon converted to a total hip arthroplasty.
|
|
The complaint is confirmed based on the customer provided photo, which displays the nail sticking out of the head of the femur.However, without return of the device for physical evaluation, the cause remains undetermined.No change in harm was identified.
|