• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RHYTHMIA HDX; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RHYTHMIA HDX; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 87037
Medical Device Problem Code Operating System Becomes Nonfunctional (2996)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported that during an ablation procedure with an intellanav stablepoint catheter, after a few maps the rhtythmia workstation was lagging, the screen was freezing.Cases were deleted in order to free some space, without success.The procedure was cancelled and there were no patient complications.The system is not expected to be returned for laboratory analysis.
 
Event or Problem Description
It was reported that during an ablation procedure with an intellanav stablepoint catheter, after a few maps the rhtythmia workstation was lagging, the screen was freezing.Cases were deleted in order to free some space, without success.The procedure was cancelled and there were no patient complications.The system is not expected to be returned for laboratory analysis.It was further reported that the rhythmia hdx software version was 4.5.
 
Additional Manufacturer Narrative
Additional information included in b5: describe event or problem it was indicated that the device will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.
 
Additional Manufacturer Narrative
The rhythmia hdx mapping system was returned to boston scientific for laboratory analysis, functional testing was conducted, the system was able to recognize known functional catheters and successfully map and perform an ablation.A 200cc map was created and no lag was visible.Hardrive was clean of extra cases.The clinical observation could not be confirmed, the device passed all relevant testing.
 
Event or Problem Description
It was reported that during an ablation procedure with an intellanav stablepoint catheter, after a few maps the rhtythmia workstation was lagging, the screen was freezing.Cases were deleted in order to free some space, without success.The procedure was cancelled and there were no patient complications.The system is not expected to be returned for laboratory analysis.It was further reported that the rhythmia hdx software version was 4.5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RHYTHMIA HDX
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112 5798
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key15310525
Report Number2124215-2022-32720
Device Sequence Number2710898
Product Code DQK
UDI-Device Identifier08714729948407
UDI-Public08714729948407
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K162793
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number87037
Device Catalogue Number87037
Device Lot Number2UA722297M
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/03/2022
Supplement Date Received by Manufacturer10/10/2022
01/25/2023
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date11/01/2022
02/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-