ZIMMER SPINE FINAL SCREWDRIVER SHAFT; INSTINCT JAVA SYSTEM Back to Search Results
Catalog Number 046W1AN00640
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a warehouse found two final drivers both with fractured tips during inspection after they were returned from a hospital.The hospital did not provide any patient or surgical information with the returned devices.This is report one of two for this event.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3003853072-2022-00007.
 
Additional Manufacturer Narrative
Corrections in d9 and h3.Additional information in h6: component, investigation type, findings, and conclusions.Inspection: visual inspection revealed that the tips of both drivers were fractured off.Dhr review the dhrs were reviewed.There are no indications of manufacturing issues which would have contributed to this event and the devices were likely conforming when they left zimvie¿s control.Potential root cause a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.This event could also be attributed to off-axis forces applied during use.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
 
Event or Problem Description
It was reported that a warehouse found two final drivers both with fractured tips during inspection after they were returned from a hospital.The hospital did not provide any patient or surgical information with the returned devices.This is report one of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FINAL SCREWDRIVER SHAFT
Common Device Name
INSTINCT JAVA SYSTEM
Manufacturer (Section D)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR  33080
Manufacturer (Section G)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR   33080
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15310867
Report Number3003853072-2022-00006
Device Sequence Number6304496
Product Code NKB
UDI-Device Identifier00889024342187
UDI-Public(01)00889024342187(10)A1432006A
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K111301
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number046W1AN00640
Device Lot NumberA1432006A
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer01/27/2023
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date01/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-