| Model Number |
SUTR ANCH MINI BC S-TAK 2.4X 8.5MM |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hand surgery the anchor broke off during insertion.The broken piece was retrieved from patient.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Upon visual inspection, it was noted that the anchor of the device was missing.The anchor was not returned for investigation.A small portion of the anchor was noted to still be assembled to the device.The cause of this remains undetermined however, probable cause can be attributed to the application of prying/leveraging forces being used.
|
| |
|