Catalog Number 2155-1
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2022
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference report 3012447612-2022-00198.
 
Event Description
It was reported that during a revision surgery to extend the construct, two incompass drivers fractured when trying to remove the previously-installed closure tops.There were no patient impacts.This is report one of two for this event.
 
Manufacturer Narrative
Corrections in h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned devices match the information in the complaint file and were examined.Visual inspection revealed a prong has fractured off the tip of both devices.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.This event could also be attributed to off-axis forces applied during use.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event Description
It was reported that during a revision surgery to extend the construct, two incompass drivers fractured when trying to remove the previously-installed closure tops.There were no patient impacts.This is report one of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL DRIVER
Type of Device
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15310926
MDR Text Key305367614
Report Number3012447612-2022-00197
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00889024398993
UDI-Public(01)00889024398993(10)62WP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132884
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2155-1
Device Lot Number62WP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/10/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-