Catalog Number MC1005T
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference report 3004788213-2022-00065.
 
Event or Problem Description
It was reported the the hook broke off an roi-c implant holder and a plate became stuck in the holder intra-operatively.There were no patient impacts.This is report one of two for this event.
 
Additional Manufacturer Narrative
Inspection: visual inspection revealed that the hook was fractured off the tip of the implant holder.Visual inspection revealed that the plate was no longer stuck in the implant holder when it was returned, but it did have a fractured tip as well.Dhr review: the dhrs were reviewed.There are no indications of manufacturing issues which would have contributed to this event and the devices were likely conforming when they left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.This event could also be attributed to off-axis forces applied during use.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
 
Event or Problem Description
It was reported the the hook broke off an roi-c implant holder and a plate became stuck in the holder intra-operatively.There were no patient impacts.This is report one of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROI-C ANCHORING PLATE
Common Device Name
ROI-C TITANIUM-COATED IMPLANT SYSTEM
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15311157
Report Number3004788213-2022-00064
Device Sequence Number20409262
Product Code OVE
UDI-Device Identifier03662663016971
UDI-Public(01)03662663016971(17)240201(10)758087
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K151934
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2024
Device Catalogue NumberMC1005T
Device Lot Number758087
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2022
Supplement Date Received by Manufacturer02/02/2023
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date02/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-