MEDOS INTERNATIONAL SÃ RL CH EXPEDIUM SPINE SYSTEM SI POLYAXIAL SCREW 5.5 X 5 X 50MM; ORTHOSIS, SPINAL PEDICLE FIXATION
|
Back to Search Results |
|
| Model Number |
179712550 |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
08/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Device report from cyprus reports an event as follows: it was reported that on (b)(6) 2022, a revision surgery was performed on a patient that had underwent lumbar spine fusion (using 8 pedicle screws) 6 months ago.One month ago, the patient had an accident and slipped and fell to the floor.After that, the patient was in terrible pain and had an x-ray and a ct scan.The pain did not go away so the doctor decided to operate on the patient again.During the revision procedure, the surgeon could see that the screws of the crosslink plate were almost unlocked and one pedicle screw was falling out, creating a black color in the field around it.When the surgeon removed the crosslink plate, the rod was cut out as well.After the rod removal, it was noticed that a screw was falling out.The surgeon used the screwdriver to unscrew it and the screw came out cut as well.The screw could not be replaced because the thread shaft of the screw was bothering to insert another screw.Finally, the surgeon decided to add two more screws at the sacrum (bigger diameter), change one screw using a bigger diameter, and to replace two rods.This report is for a expedium spine system si polyaxial screw 5.5 x 5 x 50mm.This is report 2 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional procodes: osh, mnh, kwp, kwq, nkb complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6: a manufacturing record evaluation was performed for the finished device product code: 179712550.Lot number: 283630.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 07/07/2020.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H6: health effect - clinical code.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained, that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A product investigation was completed: visual analysis of the returned sample revealed that the screw was observed to have broken from shaft, the fragment was not received in the evidence provided.No other issue was identified.A dimensional inspection was unable to be performed due to post manufacturing damage.A functional evaluation cannot be completed due to the device was received itself and wit post manufacturing damage.The complaint condition was not able to be replicated.The drawings reflecting the current and manufactured revisions were reviewed.As part of depuy synthese quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|