• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC KEEL TIB TRAY CEM SZ2; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC KEEL TIB TRAY CEM SZ2; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY Back to Search Results
Model Number 86-4180
Medical Device Problem Codes Loss of or Failure to Bond (1068); Migration (4003)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Primary pfc fb operation approx 15 years ago.Tibia needed revising.Cr femur remains insitu - tibia removed and mbt tray with sleeve and stem inserted doi: jan 2007.Dor: (b)(6) 2022.Unknown side.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: b3, d6b and h6 (removed device code implant loosening: interface - unknown (loss of or failure to bond.
 
Event or Problem Description
Additional information received: can you confirm which interface was loose? it was the cement to bone interface.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available photo evidence were reviewed and cannot be established a device movement without a progressive decay of implant position in relation to initial implantation position.Based on the information provided as evidence, a defect or malfunction cannot be determined, therefore the allegation is not confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC KEEL TIB TRAY CEM SZ2
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15312021
Report Number1818910-2022-16835
Device Sequence Number6192583
Product Code JWH
UDI-Device Identifier10603295216421
UDI-Public10603295216421
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K961685
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86-4180
Device Catalogue Number864180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer08/28/2022
11/08/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date08/31/2022
11/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC KEEL TIB TRAY CEM SZ2; SIGMA CRVD GVF INS 2 10MM; UNK KNEE FEMORAL SIGMA; UNK KNEE PATELLA SIGMA
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexFemale
-
-