| Model Number |
8400D |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was returned for evaluation.Visual inspection showed evidence of physical impact to the monitor.Functional testing duplicated the reported problem.The root cause of the reported issue was found to be a corrupted main board memory chip.The main board was replaced.The device history record (dhr) review was not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.Operator of device is unknown.No product information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the device was showing an error message "ee-prom".No patient injury was reported.
|
| |
|