Catalog Number 8065751593
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample has been received by manufacturing that has not yet been evaluated.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that a ophthalmic laser probe got caught in a port and could not be inserted into the patient's eye during surgery.The surgery was completed after replacing the product with another one.There's no patient harm.
 
Additional Manufacturer Narrative
The 25 gauge illuminated flex curved laser probe was received.The 25 gauge illuminated flex curved laser probe sample testing was performed and revealed no problem with trocar insertion.Unrelated to the reported event, the sample was found to have a non-conforming optical fiber at subminiature assembly (sma) connector, recessed.However, the sample¿s laser radio frequency identification (rfid) tag was read, and the laser probe sample was found to not be associated with the customer¿s reported lot.The laser probe sample returned is unrelated to the customer¿s reported event.Based on the information available, the customer reported event cannot be confirmed.A 25 gauge illuminated flex curved laser probe manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A non-conformance based review of the batch/lot/serial number was performed and a potential contributing factor to the reported complaint was identified.A review for complaints reported against this lot/batch/serial number was performed.Two potentially relevant complaints were found and reviewed as part of this investigation.The complaints were determined to not be relevant to this investigation.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.The customer reported event cannot be confirmed.Thus, based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LASER PROBES
Common Device Name
INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15312116
Report Number2028159-2022-01277
Device Sequence Number19519884
Product Code HQE
UDI-Device Identifier00380657515936
UDI-Public00380657515936
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K946135
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number8065751593
Device Lot Number1463EA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer10/13/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date11/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1