• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY TREVO NXT 3MM X 32MM - US; NEUROVASCULAR MECHANICAL THROMBECTOMY DEVICE FOR ACUTE ISCHEMIC STROKE TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY TREVO NXT 3MM X 32MM - US; NEUROVASCULAR MECHANICAL THROMBECTOMY DEVICE FOR ACUTE ISCHEMIC STROKE TREATMENT Back to Search Results
Model Number 90312
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hemorrhagic Stroke (4417)
Date of Event 05/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the unruptured cerebral aneurysm case, physician deployed the subject stent retriever into the torturous vessel.The subject stent retriever could not be used and was removed from the patient's anatomy.Post-procedure patient was presented with severe stroke (sah).Physician thought that the adverse event was due to the subject stent retriever deployment.Patient was given antihypertensive and hemostatic medications.One month later, patient was recovered but had sequelae.No additional information available.
 
Additional Manufacturer Narrative
There are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The patient had severe vascular tortuosity, so the subject stent retriever was deployed in a vessel and a catheter was attempted to be guided to the distal vessel using the subject stet retriever as a support, but it could not be guided and it was removed.The reported 'patient intracranial hemorrhage' is a known and anticipated complication to these types of procedures and patient condition and is listed as such in the device directions for use.Therefore, an assignable cause of anticipated procedural complication will be assigned to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREVO NXT 3MM X 32MM - US
Common Device Name
NEUROVASCULAR MECHANICAL THROMBECTOMY DEVICE FOR ACUTE ISCHEMIC STROKE TREATMENT
Manufacturer (Section D)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer (Section G)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15312461
Report Number3012931345-2022-00167
Device Sequence Number6276874
Product Code POL
UDI-Device Identifier07613327299922
UDI-Public07613327299922
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200117
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number90312
Device Catalogue Number90312
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/03/2022
Initial Report FDA Received Date08/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AXS CATALYST CATHETER(STRYKER); SURPASS FLOW DIVERTER STENT (STRYKER)
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age60 YR
Patient SexFemale
Patient Weight43 KG
Patient RaceAsian
-
-