MAUDE Adverse Event Report: SMITH & NEPHEW, INC. HEALICOIL SA PK 5.5MM W/3 UB-BL CBBL LT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. HEALICOIL SA PK 5.5MM W/3 UB-BL CBBL LT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72203380
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Rupture (2208)
Date of Event 08/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during use, two (2) healicoil suture anchor broke when being screwed in, the pieces were removed using a grasper.Surgery was resumed after a non-significant surgical delay, using a back-up device instead.It was necessary to drill a new bone hole.The current state of the patient is normal.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H10: h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.An evaluation of the customer provided image showed the anchor inside of the patient.The anchor has broken ribs protruding.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor material drawing/specifications found that certifications and testing data are required for raw material.One undated intraoperative photo was provided for review and confirms the reported breakage.Per case details, the broken pieces were retrieved from the patient.The procedure was completed using a back-up device.No further risk to the patient is anticipated as a result of the additional bone hole.Since no further harm is anticipated no further clinical assessment is warranted at this time.There was a relationship found between the device and the reported event.The complaint was confirmed and the root cause was associated with unintended use of the device.Factors that may have contributed to the reported event include not maintaining inserter alignment throughout insertion to ensure implant integrity or use of excessive force during insertion.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALICOIL SA PK 5.5MM W/3 UB-BL CBBL LT
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15312526
Report Number1219602-2022-01259
Device Sequence Number15035490
Product Code MBI
UDI-Device Identifier00885554026313
UDI-Public00885554026313
Combination Product (Y/N)N
Initial Reporter CountryHK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72203380
Device Catalogue Number72203380
Device Lot Number2080756
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer10/05/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date10/06/2022
Date Device Manufactured10/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-