MAUDE Adverse Event Report: DEPUY SPINE INC VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 45MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 45MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 186731745
Device Problem Difficult to Advance (2920)
Patient Problems Failure of Implant (1924); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 08/02/2022
Event Type  Injury  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional device product codes: mni, kwp, kwq, and mnh.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2022, during a revision for a plif procedures (l3) for treating asd.After a screw (6.0 mm in diameter) was removed from the lesion, a replacement screw was prepared.The screw did not advance further against the much-hardened bone.When the surgeon forced the screw with the screwdriver, its tip broke off.The procedure was completed less than 30-minute surgical delay.This report involves one (1) viper system cortical fix polyaxial screw 5.5 7 x 45mm.This is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d3, d9.H3, h4, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.H6: health effect - clinical code.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a manufacturing record evaluation was performed for the finished device product code: 186731745.Lot number: 338340.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 24/03/2022.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the 5.5 ti cort fix 7x45mm.A dimensional inspection was unable to be performed due to post manufacturing damage.A functional test to assess the will not seat/advance condition can not be conducted since the allegation is related to bone as part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the 5.5 ti cort fix 7x45mm was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H6: health effect - clinical code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 45MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15312796
MDR Text Key298813579
Report Number1526439-2022-01491
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034352015
UDI-Public10705034352015
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K110216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number186731745
Device Catalogue Number186731745
Device Lot Number338340
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/01/2022
Initial Date FDA Received08/29/2022
Supplement Dates Manufacturer Received09/22/2022
10/14/2022
Supplement Dates FDA Received10/11/2022
10/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
EXP A QUICKCONNECT SCREWDRIVER
Patient Outcome(s) Required Intervention;
-
-