• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 5-6 RT 12MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 5-6 RT 12MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74027254
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Fall (1848)
Date of Event 08/06/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tka, a patient fell and ruptured its closure.The insert jrny ii bcs xlpe art isrt sz 5-6 rt 12mm was replaced with treatment of antibiotics.The current health status of the patient is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, after a tka done on (b)(6)2022, a patient required a revision due to an unspecified infection.This was done on (b)(6)2022.The patient fell and ¿ruptured¿ the closure.The second replacement was done on (b)(6) 2022.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The patient's current condition is unknown, and the patient impact beyond the 3 surgeries in a short period of time, could not be determined based on the limited information provided.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed in warnings and precautions, postoperative section that use extreme care in patient handling is needed and postoperative patient care and directions and warnings to patients by physicians are extremely important.Protected weight bearing with external support is recommended for a period of time to allow healing.Normal daily activity may be resumed at the physician¿s direction.Patients should be directed to seek medical opinion before entering potentially adverse environments that could affect the performance of the implant.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury and/or postoperative care.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS XLPE ART ISRT SZ 5-6 RT 12MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15312956
Report Number1020279-2022-03866
Device Sequence Number6276580
Product Code JWH
UDI-Device Identifier00885556179512
UDI-Public00885556179512
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74027254
Device Catalogue Number74027254
Device Lot Number16LM00345
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/01/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-