| Model Number |
1012450-12 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a heavily calcified, mildly tortuous left anterior descending artery.The 3.25x12mm nc trek balloon dilatation catheter (bdc) met resistance with anatomy during advancement.The bdc ruptured during the first inflation at 12 atmospheres and during removal resistance was met with anatomy.It is unknown if another device was used to used to successfully complete the procedure.There was no adverse patient effects and no clinically significant delay.No additional information was provided.
|
| |
|