• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL LMA UNIQUE EVO SIZE 4; AIRWAY, OROPHARYNGEAL, ANESTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL LMA UNIQUE EVO SIZE 4; AIRWAY, OROPHARYNGEAL, ANESTH Back to Search Results
Model Number IPN923211
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that: "proximal end of the lma cracked.The reported issue was detected prior to use on the patient.There was no patient injury.
 
Event or Problem Description
It was reported that: "proximal end of the lma cracked.The reported issue was detected prior to use on the patient.There was no patient injury.
 
Additional Manufacturer Narrative
Qn# (b)(4).The actual device was returned and sent to the manufacturing site for investigation.The manufacturing site reports: "dhr was reviewed for this complaint.There was no abnormality found.The complaint sample was observed unloaded from the original product pouch which was not returned along with the sample upon receipt.There were no broken pieces/chips from the connector was returned for evaluation.The connector is made of polypropylene (pp) which is generally known as a strong material, which is a very durable plastic that are resistant to fatigue and has good impact strength.The device passed the necessary final inspection.The actual cause of the incident is classified as undetermined.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LMA UNIQUE EVO SIZE 4
Common Device Name
AIRWAY, OROPHARYNGEAL, ANESTH
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
THE LARYNGEAL MASK COMPANY
6 battery road #07-02
singapore 04990 9
SN   049909
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15314349
Report Number8040412-2022-00239
Device Sequence Number17522304
Product Code CAE
UDI-Device Identifier15060112318914
UDI-Public15060112318914
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIPN923211
Device Catalogue Number1D5040
Device Lot Number11F22C0272
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer09/19/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-