| Model Number |
IPN923211 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that: "proximal end of the lma cracked.The reported issue was detected prior to use on the patient.There was no patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that: "proximal end of the lma cracked.The reported issue was detected prior to use on the patient.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The actual device was returned and sent to the manufacturing site for investigation.The manufacturing site reports: "dhr was reviewed for this complaint.There was no abnormality found.The complaint sample was observed unloaded from the original product pouch which was not returned along with the sample upon receipt.There were no broken pieces/chips from the connector was returned for evaluation.The connector is made of polypropylene (pp) which is generally known as a strong material, which is a very durable plastic that are resistant to fatigue and has good impact strength.The device passed the necessary final inspection.The actual cause of the incident is classified as undetermined.".
|
| |
|