A getinge (fse) evaluated the unit.Powered device 'on' and unit successfully passed all self-tests.Connected known iab and known system trainer and initiated autofill.Unit completed autofill and continued pumping on ecg trigger at 80bpm for approximately 120 minutes.Unable to reproduce 'leak in circuit' alarm at this time.However, while the unit was pumping, it would intermittently alarm 'low vacuum'.Commenced troubleshooting and discovered average vacuum was not in specifications.Also discovered k6, k6a, k7, and k8 leak test was intermittently failing.Replaced 5000 hour pm kit and replaced drive manifold.Full pm, safety, calibration, and functionality checks passed factory specifications.Allowed unit to continue pumping on ecg trigger at 80bpm for approximately 60 minutes without any alarms or abnormal function occurring.Device is functioning in accordance with factory specifications at this time and is cleared for clinical use.
|