Model Number 466P306AU
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Internal Organ Perforation (1987); Perforation (2001); Perforation of Vessels (2135)
Date of Event 08/13/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter occupation occupation: other, senior counsel, litigation.Date of event: please note that the exact event date is unknown and the event date is the complaint awareness date.It was reported that a patient underwent placement of a trapease vena cava filter.The filter subsequently malfunctioned and caused filter perforation and tilt.The indication for the filter implant has not been provided and there is currently no additional information available for review.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease ivc filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Ivc filter tilt has been associated with practitioner technique and vessel anatomy, specifically asymmetry and tortuosity.Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu).The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.The timing and mechanism of the tilt and perforation has not been reported at this time and a clinical conclusion could not be determined as to the cause of the event.Without procedural films or post implant imaging available for review, the reported events could not be confirmed or further clarified.There is nothing to suggest that the reported event is related to the design and/or manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.Please note that this is the initial report for this product.
 
Event or Problem Description
As reported by the legal brief, the patient underwent placement of a trapease vena cava filter.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to filter perforation and tilt.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering, and other damages.
 
Event or Problem Description
Additional information received per the medical records indicate that the patient has a history of multiple recurrent deep vein thrombosis while on anticoagulants, chronic diarrhea, hypothyroidism and bipolar disorder (managed with medication).One day after the patient was admitted, the filter was placed because of the patient complained of pain.The filter was deployed at the l2 level below the renal veins.Additional information received per the patient profile form (ppf) states that the patient experienced perforation of filter struts outside the inferior vena cava and filter tilt.The patient became aware of the reported events approximately nine years and eleven months after the index procedure.The patient reported that they experienced anxiety/mental anguish, pain (stomach, lower back and leg), shortness of breath and organ perforation related to the filter.The patient also experienced urinary tract infections, intestinal problems and poly peripheral neuropathy.The patient stated that the cause of these conditions is unknown.
 
Additional Manufacturer Narrative
After further review of additional information received, the following sections have been updated accordingly: a2, b3, b4, b5, b7, d1, d4, g2, g3, g6, h1, h2, h4 and h6.The product remains implanted and unavailable for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.Additional information is pending and will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
It was reported that a patient underwent placement of a trapease vena cava filter.The filter subsequently malfunctioned and caused filter perforation and tilt.The patient reported becoming aware of inferior vena cava (ivc and organ perforation and tilt, approximately nine years and eleven months post implant.The patient also reported anxiety, stomach, lower back and leg pain, shortness of breath and organ perforation related to the filter.The patient also reported experiencing urinary tract infections and poly peripheral neuropathy of unknown origins.According to the medical record the patient history was noted for multiple recurrent deep vein thrombosis while on anticoagulants, chronic diarrhea, hypothyroidism and bipolar disorder (managed with medication).One day after the patient was admitted, the filter was placed because of the patient complained of pain.The filter was deployed at the l2 level below the renal veins.The product remains implanted and unavailable for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The trapease ivc filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Ivc filter tilt has been associated with practitioner technique and vessel anatomy, specifically asymmetry and tortuosity.Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu).The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.The timing and mechanism of the tilt and perforation has not been reported at this time and a clinical conclusion could not be determined as to the cause of the event.Without procedural films or post implant imaging available for review, the reported events could not be confirmed or further clarified.Pain, urinary tract infections, anxiety, neuropathy, and shortness of breath do not represent a device malfunction and may be related to underlying patient specific issues.There is nothing to suggest that the reported event is related to the design and/or manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEASE PVCF FEM/JUG 55CM CSI
Common Device Name
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key15315574
Report Number9616099-2022-05929
Device Sequence Number5165155
Product Code DTK
UDI-Device Identifier20705032009451
UDI-Public20705032009451
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K020316
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2013
Device Model Number466P306AU
Device Catalogue Number466P306AU
Device Lot Number15206199
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer11/03/2022
12/06/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date12/02/2022
12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/03/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age53 YR
Patient SexFemale
-
-