| Catalog Number |
14-534900 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that the prongs on the tip of a zyston inserter broke off while installing a cage intra-operatively.The prongs were recovered with a short delay, but without patient impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection revealed the prongs have fractured off of part number 14-534900.Root cause: root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that the prongs on the tip of a zyston inserter broke off while installing a cage intra-operatively.The prongs were recovered with a short delay, but without patient impacts.
|
| |
|
Search Alerts/Recalls
|