• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CEMENT DRILL 4.5MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CEMENT DRILL 4.5MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 71369045
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during a thr 2 stage revision (1 stage) procedure, while surgeon was in the process of removing cement from the femoral canal with the cement drill 4.5mm from the renovation set, the drill tip snapped off.This caused the device became lodged in the femoral canal.An attempt was made to retrieve the drill bit initially, the fluoroscopy showed the drill tip wedged between a revision knee stem and the femoral cortex.Surgeon commented that this adverse event would be addressed during the revision of the knee prosthesis by removing the broken drill tip.Surgery was resumed, with more than an one hour delay.It is unknown how surgery was finished.The current health status of the patient is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
B5 (event description updated), d9 (device not available for evaluation).
 
Event or Problem Description
It was reported that, during the first stage of a two-stage revision surgery of competitor devices, the tip of the cement drill 4.5mm snapped off while removing cement from the femoral canal and became lodged in the femoral canal.An attempt was made to retrieve the drill bit initially; the fluoroscopy showed the drill tip wedged between a revision knee stem and the femoral cortex.The surgeon commented that this adverse event would be addressed during the revision of the knee prosthesis by removing the broken drill tip.Surgery was resumed, with more than one hour delay.It is unknown how surgery was finished.The current health status of the patient is unknown.
 
Additional Manufacturer Narrative
Results of investigation: the device was not returned for evaluation but the reported event could be confirmed with the conclusions of the clinical/medical review using the fluoroscopic image.The clinical/medical investigation concluded that, based on the limited information provided, there were no clinical factors presented which would have contributed to the event.The provided fluoroscopic image was reviewed and supports the reported ¿drill tip wedged between the revision knee stem and the femoral cortex.¿ it was noted the removal of the drill bit would be addressed during the revision.It is unknown if this noted ¿revision¿ is already planned or now required due to this drill tip being left behind.No additional information was provided.The material composition is 17-4 ph 55t/ astm f899.The drill bit is not intended for long term internal exposure and long term implantation data is not available.The patient impact beyond possible micro-motion and/or migration, and local irritation/discomfort cannot be determined.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could be corroborated, as the device broke as shown in the fluoroscopic image and the issue would be addressed during the revision of the knee prosthesis by removing the broken drill tip, besides the surgery had a delay of more than one hour.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CEMENT DRILL 4.5MM
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15316108
Report Number1020279-2022-03872
Device Sequence Number5505490
Product Code KTT
UDI-Device Identifier03596010437617
UDI-Public03596010437617
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K093047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71369045
Device Catalogue Number71369045
Device Lot Number05AM01631
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer08/09/2022
10/04/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date09/27/2022
10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-