| Catalog Number |
2N8378 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a small piece broke off from the smaller hub of a clearlink system non-dehp catheter extension set.This was observed when getting ready for a surgery and the vein was being accessed.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.Visual inspection was performed which observed a damage to the standard bore two piece luer lock body.Functional testing was performed including clear passage and pressure testing; and the results were satisfactory.The reported condition was verified during visual inspection.The cause of the condition was due to a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|