• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number A2040-120
Medical Device Problem Codes Deflation Problem (1149); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the left mid superficial femoral artery (sfa) with 100% stenosis and mild calcification.The 4.0x120mm armada 14 balloon catheter was advanced to the target lesion and inflated once at 8 atmospheres (atms).Contrast mix was 70% saline and 30% contrast.During deflation, the balloon would not deflate when the inflation device pulled negative.The inflation device was switched to a 30ml syringe, negative was held for 40 seconds and the balloon was slowly deflated.The balloon was withdrawn ultimately deflated.There was no issue with the inflation device as it was used successfully throughout the procedure.There was no resistance removing the deflated balloon.There was no difficulty removing the stylet or protective sheath and the device was prepared (air aspiration) outside the anatomy prior to use.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual, dimensional, and functional analysis was performed on the returned device.The reported deflation issue was not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints reported from this lot.The investigation was unable to determine a cause for the reported deflation issue.It was reported that the contrast mix used in this event was 70% saline and 30% contrast.It should be noted that the armada 14 instruction for use (ifu) lists material required with a contrast mix of 60% contrast diluted 1:1 with normal saline.In this event, it could not be determined if the contrast/saline mix ratio caused or contributed to the difficulty.It may be possible that the distal shaft was bent or entrapped within the anatomy causing the deflation lumen to become closed off; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15316373
Report Number2024168-2022-09229
Device Sequence Number5505504
Product Code DQY
UDI-Device Identifier08717648156168
UDI-Public08717648156168
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberA2040-120
Device Catalogue NumberA2040-120
Device Lot Number1110541
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
Patient Weight79 KG
-
-